Details
Advantages
Design innovation extended from transducers to materials.
Medical dual-element and ultra-wideband transducer technology.
Micron-precision manufacturing.
Mass production and quality control capability.
Secure, controllable supply chain and cost.
Product Details
ST120
Parameter |
Transmit |
Receive (Dual-frequency) |
Type |
Convex-array |
Linear-array |
Elements |
1 (850 kHz) |
128 |
R |
R26 (850 kHz) |
L102.4 mm |
-3 dB Center Freq. |
850 kHz ± 8% (Convex I) |
850 kHz ± 8% |
-3 dB RBW |
850 kHz ± 57 kHz |
850 kHz ± 57 kHz |
TVR |
≥159 dB |
/ |
Receiving Sensitivity |
/ |
≥-212 dB |
Pitch |
/ |
0.8 mm |
Kerf |
/ |
0.1 mm |
Width |
/ |
0.7 mm |
Transverse Aperture |
16 mm (850 kHz) |
6 mm (850 kHz + 1150 kHz) |
Plug |
/ |
Customer-supplied |
CSM |
Connector Pin Map |
Preamplifier PCBA |
Prototype Materials |
/ |
/ |
SU100
Parameter |
Specification |
Type |
Linear-array (Monostatic) |
Elements |
40 |
R |
L176 mm |
-6 dB Center Freq. |
290 kHz +/- 10% |
-6 dB RBW |
≤68% |
Pitch |
4.4 mm |
Kerf |
0.3 mm |
Width |
4.1 mm |
Transverse Aperture |
17 mm |
Plug |
Customer-supplied |
CSM |
Connector Pin Map |
Prototype Materials |
/ |