External ultrasound imaging probe (ultra-wideband probe)

  • Ultra-wideband: Ultra-wideband design provides higher signal-to-noise ratio and clearer images
  • High sensitivity: higher image resolution and penetration
  • Core components are self-produced and self-developed: All core components and processes are self-developed and self-produced, which is more stable
  • High reliability: Mature process technology and quality control bring higher reliability

Details

Details

Ultra-Wideband

Number of array elements

Arc/Width

Center frequency (MHz)

Frequency (range)

(MHz)

-6dB bandwidth 

Sensitivity 

Array element spacing (mm)

Focus depth (mm)

Maximum detection depth (mm)

CWC6-1 (convex array)

192

R55

3.4

1-6

>100%

>-45dB

0.35

(Before bending)

65

240

CWL18-5 (Linear Array)

192

L38.4

15

5-18

>100%

>-55dB

0.2

15

40

CWS5-1 (Phase-controlled Array)

80

L20

3

1-5

>100%

>-55dB

0.25

80

160